Remote Commands
Commands
6
6-33
Shows PT3926 when the RTD default type is a 100-ohm RTD with curve
α=0.003926 ohms/ohm/°C.
SP_SET
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Host Serial Port Set command) Sets the RS-232-C settings for the Calibrator rear panel
SERIAL 1 FROM HOST serial port, which is kept in the Calibrator non-volatile
memory. (While saving configuration data in the non-volatile memory, a period of about
2 seconds, the Calibrator does not respond to remote commands.) (To set the parameters
for the rear panel SERIAL 2 TO UUT serial port, see the UUT_SET command.) The
factory default values are shown below in BOLD type. (To return to the factory defaults,
see the FORMAT SETUP command.)
The interface selection sets the command response, with command echo back for
commands and error messages with TERM (terminal) or no echo back with COMP
(computer).
Parameters: <baud rate value>, 300, 600, 1200, 2400, 4800, 9600
<interface>, TERM (terminal), COMP (computer)
<flow control>, XON (xon/xoff), NOSTALL (none), RTS (rts/cts)
<number data bits>, DBIT7 (7 bits) or DBIT8 (8 bits)
<number stop bits>, SBIT1 (1 bit) or SBIT2 (2 bits)
<parity>, PNONE (none), PODD (odd),PEVEN (even)
<end of line char.> CR (carriage return), LF (line feed),
CRLF (carriage return/line feed)
Example: SP_SET 9600,TERM,XON,DBIT8,SBIT1,PNONE,CRLF
Set the parameters for the rear panel SERIAL 1 FROM HOST serial port to the factory
default values.
SP_SET?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Host Serial Port Set query) Shows the RS-232-C settings for the Calibrator rear panel
SERIAL 1 FROM HOST serial port. (To return the parameters for the rear panel
SERIAL 2 TO UUT serial port, see the UUT_SET? command.) The factory default
values are shown below in BOLD type. (To return to the factory defaults, see the
FORMAT SETUP command.)
Responses: <baud rate value>, 300, 600, 1200, 2400, 4800, 9600
<interface>, TERM (terminal), COMP (computer)
<flow control>, XON (xon/xoff), NOSTALL (none), RTS (rts/cts)
<number data bits>, DBIT7 (7 bits) or DBIT8 (8 bits)
<number stop bits>, SBIT1 (1 bit) or SBIT2 (2 bits)
<parity>, PNONE (none), PODD (odd),PEVEN (even)
<end of line char.> CR (carriage return), LF (line feed),
CRLF (carriage return/line feed)
Example: SP_SET? shows 9600,TERM,XON,DBIT8,SBIT1,PNONE,CRLF
Shows the parameters for the rear panel SERIAL 1 FROM HOST serial port, as shown,
when set to the factory default values.
SPLSTR
Sequential Overlapped
Coupled
IEEE-488 RS-232
x xx
x
x
(Serial Poll String command) Sets the Serial Poll String (string up to 40 characters)
which is kept in the Calibrator non-volatile memory. (While saving configuration data in
the non-volatile memory, a period of about 2 seconds, the Calibrator does not respond to