5502A
Operators Manual
7-10
Table 7-5. Verification Tests for DC Current (AUX) (cont.)
Range Output Lower Limit Upper Limit
2.99999 A 1.09000 A 1.08979 A 1.09021 A
2.99999 A -1.09000 A -1.09021 A -1.08962 A
2.99999 A 2.99000 A 2.98906 A 2.99094 A
2.99999 A -2.99000 A -2.99094 A -2.98906 A
20.5000 A 0.0000 A -0.0005 A 0.0005 A
20.5000 A 11.0000 A 10.9953 A 11.0046 A
20.5000 A -11.0000 A -11.0046 A 10.9953 A
20.5000 A 20.0000 A 19.9833 A 20.0168 A
20.5000 A -20.0000 A -20.0168 A -19.9833 A
Table 7-6. Verification Tests for Resistance
Range Output Lower Limit Upper Limit
10.999 Ω 0.000 Ω -0.0010 Ω 0.0010 Ω
10.999 Ω 2.000 Ω 1.9989 Ω 2.0011 Ω
10.999 Ω 10.900 Ω 10.8980 Ω 10.9019 Ω
32.999 Ω 11.900 Ω 11.8974 Ω 11.9025 Ω
32.999 Ω 19.000 Ω 18.9967 Ω 19.0032 Ω
32.999 Ω 30.000 Ω 29.9958 Ω 30.0042 Ω
109.999 Ω 33.000 Ω 32.9962 Ω 33.0037 Ω
109.999 Ω 109.000 Ω 108.9909 Ω 109.0090 Ω
329.999 Ω 119.000 Ω 118.9896 Ω 119.0103 Ω
329.999 Ω 190.000 Ω 189.9847 Ω 190.0153 Ω
329.999 Ω 300.000 Ω 299.9770 Ω 300.0230 Ω
1.09999 kΩ 0.33000 kΩ 0.329749 kΩ 0.330251 kΩ
1.09999 kΩ 1.09000 kΩ 1.089921 kΩ 1.090078 kΩ
3.29999 kΩ 1.19000 kΩ 1.189896 kΩ 1.190103 kΩ
3.29999 kΩ 1.9000 kΩ 1.899847 kΩ 1.900153 kΩ
3.29999 kΩ 3.00000 kΩ 2.999770 kΩ 3.000230 kΩ
10.9999 kΩ 3.3000 kΩ 3.29974 kΩ 3.30025 kΩ
10.9999 kΩ 10.9000 kΩ 10.89921 kΩ 10.90078 kΩ
32.9999 kΩ 11.9000 kΩ 11.89896 kΩ 11.90103 kΩ
32.9999 kΩ 19.0000 kΩ 18.99847 kΩ 19.00153 kΩ
32.9999 kΩ 30.0000 kΩ 29.99977 kΩ 30.00230 kΩ