Maintenance
Performance Tests
7
7-11
Table 7-6. Verification Tests for Resistance (cont.)
Range Output Lower Limit Upper Limit
109.999 kΩ 33.000 kΩ 32.9971 kΩ 33.0028 kΩ
109.999 kΩ 109.000 kΩ 108.9910 kΩ 109.0089 kΩ
329.999 kΩ 119.000 kΩ 118.9872 kΩ 119.0127 kΩ
329.999 kΩ 190.000 kΩ 189.9809 kΩ 190.0191 kΩ
329.999 kΩ 300.000 kΩ 299.9710 kΩ 300.0290 kΩ
1.09999 MΩ 0.33000 MΩ 0.329961 MΩ 0.330038 MΩ
1.09999 MΩ 1.09000 MΩ 1.089878 MΩ 1.090121 MΩ
3.29999 MΩ 1.19000 MΩ 1.189839 MΩ 1.190160 MΩ
3.29999 MΩ 1.90000 MΩ 1.899761 MΩ 1.900239 MΩ
3.29999 MΩ 3.00000 MΩ 2.999640 MΩ 3.000360 MΩ
10.9999 MΩ 3.3000 MΩ 3.29846 MΩ 3.30153 MΩ
10.9999 MΩ 10.9000 MΩ 10.89504 MΩ 10.90495 MΩ
32.9999 MΩ 11.9000 MΩ 11.88857 MΩ 11.91142 MΩ
32.9999 MΩ 19.0000 MΩ 18.98325 MΩ 19.01675 MΩ
32.9999 MΩ 30.0000 MΩ 29.99750 MΩ 30.02500 MΩ
109.999 MΩ 33.000 MΩ 32.8650 MΩ 33.1350 MΩ
109.999 MΩ 109.000 MΩ 108.5610 MΩ 109.4390 MΩ
329.999 MΩ 119.000 MΩ 118.4240 MΩ 119.5760 MΩ
329.999 MΩ 290.000 MΩ 288.7400 MΩ 291.2600 MΩ
1100.00 MΩ 400.00 MΩ 394.700 MΩ 405.300 MΩ
1100.00 MΩ 640.00 MΩ 631.820 MΩ 648.180 MΩ
1100.00 MΩ 1090.00 MΩ 1076.420 MΩ 1103.580 MΩ
Table 7-7. Verification Tests for AC Voltage (Normal)
Range Output Frequency Lower Limit Upper Limit
32.999 mV 3.000 mV 45 Hz 2.977 mV 3.022 mV
32.999 mV 3.000 mV 10 kHz 2.977 mV 3.022 mV
32.999 mV 30.000 mV 9.5 Hz 28.350 mV 31.650 mV,
32.999 mV 30.000 mV 10 Hz 29.944 mV 30.056 mV
32.999 mV 30.000 mV 45 Hz 29.956 mV 30.044 mV
32.999 mV 30.000 mV 1 kHz 29.956 mV 30.044 mV
32.999 mV 30.000 mV 10 kHz 29.956 mV 30.044 mV
32.999 mV 30.000 mV 20 kHz 29.944 mV 30.056 mV