5502A
Operators Manual
7-18
Table 7-11. Verification Tests for Thermocouple Simulation
TC Type Output, °C Lower Limit, mV Upper Limit, mV
10 μV/°C
0.00 °C (0.0000 mV) -0.0030 0.0030
100.00 °C (1.0000 mV) 0.99696 1.00304
-100.00 °C (-1.0000 mV) -1.00304 -0.99696
1000.00 °C (10.0000 mV) 9.99660 10.00340
-1000.00 °C (10.0000 mV) -10.0034 -9.9966
10000.00 °C (100.0000 mV) 99.9930 100.0070
-10000.00 °C (-100.0000 mV) -100.0070 -99.9930
Table 7-12. Verification Tests for Thermocouple Measurement
TC Type Input, mV Lower Limit, °C Upper Limit, °C
10 μV/°C
0.00 °C (0,0000 mV) -0.30 -0.30
10000.00 °C (100.0000 mV) 9999.30 10000.70
-10000.00 °C (-100.0000 mV) -10000.70 -9999.30
30000.00 °C (300.0000 mV) 29998.50 30001.50
-30000.00 °C (-300.0000 mV) -30001.50 -29998.50