Performance Tests
7
7-19
Table 7-13. Verification Tests for Phase Accuracy, V and V
Range,
Normal
Output, V
Output,
Normal V
Frequency
Range,
AUX
Output
Output,
AUX
Phase °
Lower
Limit°
Upper
Limit °
3.29999 3.00000
65 Hz
3.29999 V 3.00000 V
0
-0.150 0.150
400 Hz -0.900 0.900
1 kHz -2.000 2.000
5 kHz -6.000 6.000
10 kHz -10.000 10.000
30 kHz -15.000 15.000
65 Hz
60
59.850 60.150
400 Hz 59.100 60.900
1 kHz 58.000 62.000
5 kHz 54.000 66.000
10 kHz 50.000 70.000
30 kHz 45.000 75.000
65 Hz
90
89.850 90.150
400 Hz 89.100 90.900
1 kHz 88.000 92.000
5 kHz 84.000 96.000
10 kHz 80.000 100.000
30 kHz 75.000 105.000
32.9999 30.0000 65 Hz 89.85 90.15
329.999 50.000 65 Hz 89.85 90.15