5502A
Operators Manual
7-20
Table 7-14. Verification Tests for Phase Accuracy, V and I
Range,
Normal
Output, V
Output,
Normal V
Frequency
Range,
AUX
Output
Output,
AUX
Phase °
Lower
Limit °
Upper
Limit °
329.999 mV
30.000 mV
65 Hz 329.99 mA 300.00 mA
0
-0.15 0.15
1 kHz 329.99 mA 300.00 mA -2.00 2.00
30 kHz 329.99 mA 300.00 mA -15.00 15.00
200.000 mV 65 Hz 2099999 A 2.00000 A -0.15 0.15
50.000 mV
65 Hz 20.5000 A 5.0000 A -0.15 0.15
400 Hz 20.5000 A 5.0000 A -0.90 0.90
30.000 mV 65 Hz 329.99 mA 300.00 mA
60
59.85 60.15
200.000 mV
65 Hz 2.99999 A 2.00000 A 59.85 60.15
65 Hz 20.5000 A 20.0000 A 59.85 60.15
400 Hz 20.5000 A 20.0000 A 59.10 60.90
32.999 mV 3.3000 V
65 Hz 329.99 mA 300.00 mA
0
-0.15 0.15
65 Hz 2.99999 A 2.00000 A -0.15 0.15
65 Hz 20.5000 A 5.0000 A -0.15 0.15
400 Hz 20.5000 A 5.0000 A -0.90 0.90
65 Hz 329.99 mA 300.00 mA
90
89.85 90.15
65 Hz 2.99999 A 2.00000 A 89.85 90.15
65 Hz 20.5000 A 20.0000 A 89.85 90.15
400 Hz 20.5000 A 20.0000 A 89.10 90.90
329.999 V 33.000 V
65 Hz 329.99 mA 300.00 mA
0
-0.15 0.15
65 Hz 2.99999 A 2.00000 A -0.15 0.15
65 Hz 20.5000 A 5.0000 A -0.15 0.15
400 Hz 20.5000 A 5.0000 A -0.90 0.90
65 Hz 329.99 mA 300.00 mA
90
89.85 90.15
65 Hz 2.99999 A 2.00000 A 89.85 90.15
65 Hz 20.5000 A 20.0000 A 89.85 90.15
400 Hz 20.5000 A 20.0000 A 89.10 90.90