SC300 Oscilloscope Calibration Option
Verification Tables 10
10-35
Table 10-9. Leveled Sine Wave Function Verification: Flatness (cont.)
Nominal Value (p-p) Frequency Measured Value (p-p)
Deviation
(mV)
1-Year Spec.
10.0 mV 160 MHz 0.30 mV
10.0 mV 200 MHz 0.30 mV
10.0 mV 220 MHz 0.30 mV
10.0 mV 235 MHz 0.30 mV
10.0 mV 250 MHz 0.30 mV
40.0 mV 500 kHz 0.70 mV
40.0 mV 1 MHz 0.70 mV
40.0 mV 1 MHz 0.70 mV
40.0 mV 2 MHz 0.70 mV
40.0 mV 5 MHz 0.70 mV
40.0 mV 10 MHz 0.70 mV
40.0 mV 20 MHz 0.70 mV
40.0 mV 50 MHz 0.70 mV
40.0 mV 100 MHz 0.70 mV
40.0 mV 125 MHz 0.90 mV
40.0 mV 160 MHz 0.90 mV
40.0 mV 200 MHz 0.90 mV
40.0 mV 220 MHz 0.90 mV
40.0 mV 235 MHz 0.90 mV
40.0 mV 250 MHz 0.90 mV
100.0 mV 500 kHz 1.60 mV
100.0 mV 1 MHz 1.60 mV
100.0 mV 1 MHz 1.60 mV
100.0 mV 2 MHz 1.60 mV
100.0 mV 5 MHz 1.60 mV
100.0 mV 10 MHz 1.60 mV
100.0 mV 20 MHz 1.60 mV
100.0 mV 50 MHz 1.60 mV
100.0 mV 100 MHz 1.60 mV
100.0 mV 125 MHz 2.10 mV
100.0 mV 160 MHz 2.10 mV
100.0 mV 200 MHz 2.10 mV