Table 10-9. Leveled Sine Wave Function Verification: Flatness (cont.)
Nominal Value (p-p) Frequency Measured Value (p-p)
Deviation
(mV)
1-Year Spec.
100.0 mV 220 MHz 2.10 mV
100.0 mV 235 MHz 2.10 mV
100.0 mV 250 MHz 2.10 mV
400.0 mV 500 kHz 6.10 mV
400.0 mV 1 MHz 6.10 mV
400.0 mV 1 MHz 6.10 mV
400.0 mV 2 MHz 6.10 mV
400.0 mV 5 MHz 6.10 mV
400.0 mV 10 MHz 6.10 mV
400.0 mV 20 MHz 6.10 mV
400.0 mV 50 MHz 6.10 mV
400.0 mV 100 MHz 6.10 mV
400.0 mV 125 MHz 8.10 mV
400.0 mV 160 MHz 8.10 mV
400.0 mV 200 MHz 8.10 mV
400.0 mV 220 MHz 8.10 mV
400.0 mV 235 MHz 8.10 mV
400.0 mV 250 MHz 8.10 mV
1.3 V 500 kHz 19.60 mV
1.3 V 1 MHz 19.60 mV
1.3 V 1 MHz 19.60 mV
1.3 V 2 MHz 19.60 mV
1.3 V 5 MHz 19.60 mV
1.3 V 10 MHz 19.60 mV
1.3 V 20 MHz 19.60 mV
1.3 V 50 MHz 19.60 mV
1.3 V 100 MHz 19.60 mV
1.3 V 125 MHz 26.10 mV
1.3 V 160 MHz 26.10 mV
1.3 V 200 MHz 26.10 mV
1.3 V 220 MHz 26.10 mV
1.3 V 235 MHz 26.10 mV