EasyManua.ls Logo

Fluke 5502A - Page 61

Fluke 5502A
378 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Features
Softkey Menu Trees 3
3-9
Table 3-1. Front-Panel Features (cont.)
b
The TC (Thermocouple) minijackis used for thermocouple simulation during thermometer
calibration, and thermocouple measurements. You must use the correct thermocouple wire and
plug when you use this connector. For example, if you will simulate a type-K thermocouple, use
type-K thermocouple wire and type-K plug to make connections.
c
The 20A terminal is the source of current output when the 20 A range is selected (3 A to 20 A).
d
The AUX (Auxiliary Output) terminals are used for ac and dc current outputs, the second voltage
output in dual-voltage modes, and ohms sense for 2-wire and 4-wire compensated resistance and
capacitance measurements, and RTD simulation.
e
The NORMAL (Normal Output) terminals are used to source ac and dc voltage, ohms, and
capacitance. The terminals are also used for Resistance Temperature Detector (RTD) simulation.

Table of Contents