5502A
Operators Manual
3-10
CALIBRATION
ENABLE
SERIAL 1
FROM HOST
IEEE-488
IN OUT
/63Hz 47Hz
600VA MAX
FLUKE CORPORATION
EVERETT WA, USA
NORMAL
WARNING:
TO AVOID ELECTRIC SHOCK GROUNDING
CONNECTOR IN POWER CORD MUST BE CONNECTED
FUSE
CAUTION
FOR FIRE PROTECTION REPLACE ONLY
WITH A 250V FUSE OF INDICATED RATING
MAINS SUPPLY
100V/120V
220V/240V
T5.0A 250V
(SB)
T2.5A 250V
(SB)
CHASSIS
GROUND
SERIAL 2
TO UUT
10MHZ
MAX
5V
PK - PK
WARNING:
TO AVOID PHYSICAL INJURY, INSURE THAT THE FILTER
IS PROPERLY INSTALLED BEFORE ENERGIZING INSTRUMENT
TO CLEAN THE FILTER:
-UNPLUG INSTRUMENT
-REMOVE FILTER
-FLUSH WITH SOAPY WATER
-DRY BEFORE REINSTALLATION
INSTALLED
OPTIONS
- SC600
- SC1100
- PQ
LR65268C
(LEM CERTIFIED)
NO INTERNAL USER SERVICEABLE
PARTS. REFER SERVICE TO
QUALIFIED SERVICE PERSONNEL
1 2
5
67
3 4
gjh011.eps
Figure 3-2. Rear-Panel Features
Table 3-2. Rear-Panel Features
The Fan Filtercovers the air intake to keep dust and contamination out of the chassis air baffles.
The fan supplies a constant cool air flow to the inner chassis. Instructions for fan filter maintenance
are in Chapter 7.
The CALIBRATION NORMAL/ENABLE slide switch is used to write enable and disable the
nonvolatile memory that stores calibration constants. Change to ENABLE to let changes be written
into memory, and change to NORMAL to protect data in memory from being overwritten. The
switch is recessed to let it to be covered with a calibration sticker to guarantee calibration integrity.
The SERIAL 2 TO UUT connector is used to transmit and receive RS-232 serial data between the
Calibrator and a UUT. Chapter 6, “Remote Commands” tells you how to use the RS-232 serial
interface for UUT communications.
The SERIAL 1 FROM HOST connector is for remote control of the Calibrator and for transmitting
internal-constant RS-232 serial data to a printer, monitor, or host computer. Chapter 5, “Remote
Operation” tells you how to use the RS-232 serial interface for remote control.
The IEEE-488 connectorError! Bookmark not defined. is a standard parallel interface for
operating the Calibrator in remote control as a Talker/Listener on the IEEE-488 bus. Refer to
Chapter 5, “Remote Operation” for bus connection and remote programming instructions.