Table 3-27. Verification Tests for AC Current (cont)
Range Output Frequency Lower Limit Upper Limit
20.5000 A 10.9000 A 45 Hz 10.8926 A 10.9075 A
20.5000 A 10.9000 A 65 Hz 10.8926 A 10.9075 A
20.5000 A 10.9000 A 500 Hz 10.8893 A 10.9107 A
20.5000 A 10.9000 A 1 kHz 10.8893 A 10.9107 A
20.5000 A 10.9000 A 5 kHz 10.6255 A 11.1745 A
20.5000 A 20.0000 A 45 Hz 19.9750 A 20.0250 A
20.5000 A 20.0000 A 65 Hz 19.9750 A 20.0250 A
20.5000 A 20.0000 A 500 Hz 19.9690 A 20.0310 A
20.5000 A 20.0000 A 1 kHz 19.9690 A 20.0310 A
20.5000 A 20.0000 A 5 kHz 19.4950 A 20.5050 A
3-26. Verifying Capacitance
Verify that the 5520A performance is within the limits in Table 3-28. Use the PM 6304C
RCL Meter directly for capacitance values up to and including 109.000 µF. Above
109.000 µF, you must use a timed charge up routine with a constant current source in
order to achieve the required test uncertainty ratio.
To verify capacitance greater than 109.000 µF, see the section titled “200 µF to 110 mF
Capacitance Verification” found later in this chapter.
Table 3-28. Verification Tests for Capacitance
Range Output
Test Frequency or
Current
Lower Limit Upper Limit
0.3999 nF 0.1900 nF 5 kHz 0.1793 nF 0.2007 nF
0.3999 nF 0.3500 nF 1 kHz 0.3387 nF 0.3613 nF
1.0999 nF 0.4800 nF 1 kHz 0.4682 nF 0.4918 nF
1.0999 nF 0.6000 nF 1 kHz 0.5877 nF 0.6123 nF
1.0999 nF 1.0000 nF 1 kHz 0.9862 nF 1.0138 nF
3.2999 nF 2.0000 nF 1 kHz 1.9824 nF 2.0176 nF
10.9999 nF 7.0000 nF 1 kHz 6.9767 nF 7.0233 nF
10.9999 nF 10.9000 nF 1 kHz 10.8693 nF 10.9307 nF
32.9999 nF 20.0000 nF 1 kHz 19.8620 nF 20.1380 nF
109.999 nF 70.000 nF 1 kHz 69.767 nF 70.233 nF
109.999 nF 109.000 nF 1 kHz 108.693 nF 109.307 nF
329.999 nF 200.000 nF 1 kHz 199.320 nF 200.680 nF
329.999 nF 300.000 nF 1 kHz 299.130 nF 300.870 nF