1-12. Resistance Specifications
Range
Absolute Uncertainty, tcal ± 5 °C
± (ppm of output + floor) [2]
Resolution Allowable
[1]
ppm of output
Floor
Time & temp since ohms zero
cal
Current [3]
90 days 1 year 12 hrs ± 1 °C 7 days ± 5 °C
0 to
10.9999
35 40 0.001 0. 01 0.0001 1 mA to 125 mA
11 to
32.9999
25 30 0.0015 0.015 0.0001 1 mA to 125 mA
33 to
109.9999
22 28 0.0014 0.015 0.0001 1 mA to 70 mA
110 to
329.9999
22 28 0.002 0.02 0.0001 1 mA to 40 mA
330 to
1.099999 k
22 28 0.002 0.02 0.001 1 mA to 18 mA
1.1 k to
3.299999 k
22 28 0.02 0.2 0.001 100 µA to 5 mA
3.3 kto
10.99999 k
22 28 0.02 0.1 0.01 100 µA to 1.8 mA
11 kto
32.99999 k
22 28 0.2 1 0.01 10 µA to 0.5 mA
33 kto
109.9999 k
22 28 0.2 1 0. 1 10 µA to 0.18 mA
110 kto
329.9999 k
25 32 2 10 0.1 1 µA to 0.05 mA
330 k to
1.099999 M
25 32 2 10 1 1 µA to 0.018 mA
1.1 Mto
3.299999 M
40 60 30 150 1 250 nA to 5 µA
3.3 Mto
10.99999 M
110 130 50 250 10 250 nA to 1.8 µA
11 Mto
32.99999 M
200 250 2500 2500 10 25 nA to 500 nA
33 Mto
109.9999 M
400 500 3000 3000 100 25 nA to 180 nA
110 Mto
329.9999 M
2500 3000 100000 100000 1000 2.5 nA to 50 nA
330 Mto
1100 M
12000 15000 500000 500000 10000 1 nA to 13 nA
[1] Continuously variable from 0 to 1.1 G .
[2] Applies for a 4-WIRE compensation only. For 2-WIRE and 2-WIRE COMP, add 5 µV per Amp of stimulus current
to the floor specification. For example, in 2-WIRE mode, at 1 k
e
, the floor specification within 12 hours of an ohms
zero cal for a measurement current of 1 mA is:
0.002
e
+ 5 µV/1 ma = (0.002 + 0.005)
e
= 0.007
e
[3] For currents lower than shown, the floor adder increases by:
Floor (new) = Floor (old) XImin/Iactual.
For example, a 50 µA stimulus measuring 100
e
, has a floor specification of: 0.0014
e
X 1 mA/50 µA = 0.028
e
,
assuming an ohms zero cal within 12 hours.