5520A
Service Manual
3-52
3-30. Verifying Thermocouple Simulation (Sourcing)
Verify that the 5520A performance is within the limits in Table 3-30. Use the HP3458A
DMM as the measurement device. Use copper connectors and copper wires.
Table 3-30. Verification Tests for Thermocouple Simulation
TC Type Output, °C Lower Limit, mV Upper Limit, mV
10 µV/°C 0.00 °C (0.0000 mV) -0.0030 0.0030
100.00 °C (1.0000 mV) 0.99696 1.00304
-100.00 °C (-1.0000 mV) -1.00304 -.99696
1000.00 °C (10.0000 mV) 9.99660 10.00340
-1000.00 °C (-10.0000 mV) -10.0034 -9.9966
10000.00 °C (100.0000 mV) 99.9930 100.0070
-10000.00 °C (-100.0000 mV) -100.0070 -99.9930
3-31. Verifying Thermocouple Measurement
Verify that the 5520A performance is within the limits in Table 3-31. Use a Fluke 5500A
Calibrator or similar instrument as the millivolt source, connected in parallel with an
HP3458A DMM. At each verification point, use the 5500A error mode controls to adjust
the calibrator output for a nominal reading on the DMM. Use copper connectors and
copper wires.
Table 3-31. Verification Tests for Thermocouple Measurement
TC Type Input, mV Lower Limit, °C Upper Limit, °C
10 µV/°C 0.00 °C (0.0000 mV) -0.30 0.30
10000.00 °C (100.0000 mV) 9999.30 10000.70
-10000.00 °C (-100.0000 mV) -10000.70 -9999.30
30000.00 °C (300.0000 mV) 29998.50 30001.50
-30000.00 °C (-300.0000 mV) -30001.50 -29998.50