Calibration and Verification
Performance Verification Tests
3
3-53
3-32. Verifying Phase Accuracy, Volts and AUX Volts
Verify that the 5520A performance is within the limits in Table 3-32, using a precision
phase meter see Figure 3-15.
Table 3-32. Verification Tests for Phase Accuracy, V and V
Range, Normal
Output, V
Output,
Normal
V
Frequency
Range, AUX
Output
Output, AUX Phase
°
Lower Limit
°
Upper Limit
°
3.29999 3.00000 65 Hz 3.29999 V 3.00000 V 0 -0.10 0.10
400 Hz -0.25 0.25
1 kHz -0.50 0.50
5 kHz -2.50 2.50
10 kHz -5.00 5.00
30 kHz -10.00 10.00
65 Hz 60 59.90 60.10
400 Hz 59.75 60.25
1 kHz 59.50 60.50
5 kHz 57.50 62.50
10 kHz 55.00 65.00
30 kHz 50.00 70.00
65 Hz 90 89.90 90.10
400 Hz 89.75 90.25
1 kHz 89.50 90.50
5 kHz 87.50 92.50
10 kHz 85.00 95.00
30 kHz 80.00 100.00
32.9999 30.0000 65 Hz 89.90 90.10
329.999 50.000 65 Hz 89.90 90.10