5520A
Service Manual
3-20
# Retrieve AC-DC difference from data file named "A40-10mA"
1.059 DOS get_acdc A40-10mA
1.060 JMPT 1.064
1.061 OPBR An error occurred during get_acdc
1.061 OPBR Press YES to try again or NO to terminate.
1.062 JMPT 1.059
1.063 JMP 1.231
# Correct the calculated value of AC Current by adding the AC-DC difference
# of the A40-series shunt used at the frequency under test
1.064 MATH MEM = (M[21] * MEM) + M[21]
# Store corrected value into the UUT
1.065 IEEE CAL_NEXT [MEM]; *OPC?[I][GTL]
1.066 IEEE ERR?[I$][GTL]
1.067 MATH MEM1 = FLD(MEM2,1,",")
1.068 JMPT 1.231
# ’Ask’ UUT for next value to calibrate
1.069 IEEE CAL_REF?[I][GTL]
Figure 3-9. Sample MET/CAL Program (cont)
3-11. DC Volts Calibration (AUX Output)
To calibrate the auxiliary dc voltage function, use the same technique as previously
described for the normal dc voltage output, except use the AUX HI and LO terminals on
the UUT. Table 3-12 lists the calibration steps for AUX dc volts.
Table 3-12. Calibration Steps for AUX DC Volts
Step 5520A Output (AUX)
1 300.000 mV
2 3.00000 V
3 7.00000 V
3-12. AC Volts Calibration (AUX Output)
To calibrate the auxiliary ac voltage function, use the same technique as previously
described for the normal ac voltage output, except use the AUX HI and LO terminals on
the UUT. Table 3-13 lists the calibration steps for AUX dc volts.
Table 3-13. Calibration Steps for AUX Output AC Volts
5520A Output (AUX)
Step
Amplitude Frequency
1 300.000 mV 100 Hz
2 300.000 mV 5 kHz
3 3.00000 V 100 Hz
4 3.00000 V 5 kHz
5 5.0000 V 100 Hz
6 5.0000 V 5 kHz
7 3.0 V 9.99 Hz