SC600 Option
Verification
6
6-49
PM 6680 Settings
PM 6680
Reading
1
PM 6680 Reading
Calibrator
Mainframe
Period
Channel Filter (Frequency) (Period)
Tolerance
5 s A On 0.3489454 s
2 s A On 0.0582996 s
50.0 ms A Off 3.872E-05 s
20.0 ms A Off 5E-08 s
10.0 ms A Off 2.5E-08 s
100 ns A Off 2.5E-13 s
50.0 ns A Off 1.25E-13 s
20.0 ns A Off 5E-14 s
10.0 ns A Off 2.5E-14 s
5.00 ns A Off 1.25E-14 s
2.00 ns C Off 5E-15 s
6-66. Wave Generator Verification
This procedure uses the following equipment:
5790A AC Measurement Standard
BNC(f) to Double Banana adapter
50 feedthrough termination
BNC cable supplied with the Calibrator Mainframe
SC600
Cable
5520A-SC600
50
Feed Through
Termination
BNC (F) to
Double Banana
Adapter
HI
LO
TRIG
GUARD
TC
20A
1000V
RMS
MAX
20V
RMS
MAX
20V
PK
MAX
1V PK
MAX
SHELLS
NOT
GROUNDED
NORMAL AUX
SCOPE
OUT
V, , ,
RTD
A, -SENSE,
AUX V
150V
PK
MAX
20V
RMS
MAX
20V PK MAX
20V PK MAX
5520A
CALIBRATOR
yg060f.eps
Figure 6-13. Wave Generator Verification Setup
For wave generation verification procedures, refer to Figure 6-13 for the proper setup
connections.
Set the Calibrator Mainframe to SCOPE mode, with the Wavegen menu on the display.
Press O on the Calibrator Mainframe to activate the output. Set the offset to 0 mV,