SC300 Option
Verification
6
6-97
Table 6-53. Leveled Sine Wave Frequency Verification
PM 6680 Settings
Calibrator Mainframe
Frequency
(Output @ 5.5 V p-p)
Channel Filter
PM 6680 Reading
(Frequency)
Tolerance
50 kHz A On 1.25 Hz
500 kHz A Off 12.5 Hz
5 MHz A Off 125.0 Hz
50 MHz A Off 1250 Hz
500 MHz C Off 12500 Hz
6-126. Leveled Sine Wave Harmonics Verification
This procedure uses the following equipment:
Hewlett-Packard 8590A Spectrum Analyzer
BNC(f) to Type N(m) adapter
BNC cable supplied with the SC300
Refer to Figure 6-24 for proper setup connections.
HP 8590A
SC300
Cable
BNC(F)
to Type N (M)
Adapter
5520A-SC300
HI
LO
TRIG
GUARD
TC
20A
1000V
RMS
MAX
20V
RMS
MAX
20V
PK
MAX
1V PK
MAX
SHELLS
NOT
GROUNDED
NORMAL AUX
SCOPE
OUT
V, , ,
RTD
A, -SENSE,
AUX V
150V
PK
MAX
20V
RMS
MAX
20V PK MAX
20V PK MAX
5520A
CALIBRATOR
yg125f.eps
Figure 6-24. Leveled Sine Wave Harmonics Verification Setup
Set the Calibrator Mainframe to SCOPE mode, with the Levsine menu on the display.
Then follow these steps to verify the leveled sine wave harmonics.
1. Using the BNC cable and BNC(f) to Type N(m) adapter, connect the SCOPE
connector on the Calibrator Mainframe to the HP 8590A.
2. Program the Calibrator Mainframe to 5.5 V p-p at each frequency listed in Table 6-
54. Press O on the Calibrator Mainframe to activate the output.