6-6
6-3. SC600 Specifications
These specifications apply only to the SC600 Option. General specifications that apply to
the Calibrator Mainframe (hereafter termed the Calibrator) can be found in Chapter 1.
The specifications are valid under the following conditions:
The Calibrator is operated under the conditions specified in Chapter 1.
The Calibrator has completed a warm-up period of at least twice the length of time
the calibrator was powered off, up to a maximum of 30 minutes.
The SC600 Option has been active longer than 5 minutes.
6-4. Volt Specifications
Table 6-1. Volt Specifications
Volt Function
dc Signal Square Wave Signal [1]
50 Load 1 M Load 50 Load 1 M Load
Amplitude Characteristics
Range
0 V to ±6.6 V 0 V to ±130 V ±1 mV to
±6.6 V p-p
±1 mV to
±130 V p-p
Resolution
Range
1 mV to 24.999 mV
25 mV to 109.99 mV
110 mV to 2.1999 V
2.2 V to 10.999 V
11 V to 130 V
Resolution
1 µV
10 µV
100 µV
1 mV
10 mV
Adjustment Range Continuously adjustable
1-Year Absolute Uncertainty,
tcal ± 5 °C
± (0.25% of
output +
40 µV)
± (0.05% of
output +
40 µV)
± (0.25% of
output + 40
µV)
± (0.1% of
output +
40 µV) [2]
Sequence 1-2-5 (e.g., 10 mV, 20 mV, 50 mV)
Square Wave Frequency Characteristics
Range 10 Hz to 10 kHz
1-Year Absolute Uncertainty,
tcal ± 5 °C
± (2.5 ppm of setting)
Typical Aberration
within 4 µs from 50% of
leading/trailing edge
< (0.5% of output + 100 µV)
[1] Selectable positive or negative, zero referenced square wave.
[2] For square wave frequencies above 1 kHz, ± (0.25% of output + 40 µV).