EasyManua.ls Logo

Fluke 5520A - AC Current, Squarewave Characteristics (Typical); AC Current, Trianglewave Characteristics (Typical)

Fluke 5520A
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Introduction and Specifications
Additional Specifications
1
1-31
1-33. AC Current, Square Wave Characteristics (typical)
Range LCOMP Risetime Settling Time Overshoot
I < 6A @ 400 Hz off 25 µs 40 µs to 1% of final value < 10% for < 1 V
Compliance
3A & 20A
Ranges
on 100 µs 200 µs to 1% of final value < 10% for < 1 V
Compliance
1-34. AC Current, Triangle Wave Characteristics (typical)
Linearity to 400 Hz Aberrations
0.3% of p-p value, from 10% to 90% point < 1% of p-p value, with amplitude > 50% of range

Table of Contents

Other manuals for Fluke 5520A

Related product manuals