Introduction and Specifications
Specifications
1
1-19
1-18. DC Power Specification Summary
Current Range
Voltage Range
0.33 mA to
329.99 mA
0.33 A to
2.9999 A
3 A to
20.5 A
Absolute Uncertainty, tcal ± 5 °C, ± (% of watts output) [1]
90 days 33 mV to 1020 V 0.021 0.019 [2] 0.06 [2]
1 year 33 mV to 1020 V 0.023 0.022 [2] 0.07 [2]
[1] To determine dc power uncertainty with more precision, see the individual “AC Voltage Specifications,” “AC
Current Specifications,” and “Calculating Power Uncertainty.”
[2] Add 0.02% unless a settling time of 30 seconds is allowed for output currents > 10 A or for currents on the highest
two current ranges within 30 seconds of an output current > 10 A.
1-19. AC Power (45 Hz to 65 Hz) Specification Summary, PF=1
Current Range
Voltage Range
3.3 mA to
8.999 mA
9 mA to
32.999 mA
33 mA to
89.99 mA
90 mA
to 329.99 mA
Absolute Uncertainty, tcal ± 5 °C, ± (% of watts output) [1]
90 days 33 to 329.999 mV 0.13 0.09 0.13 0.09
330 mV to 1020 V 0.11 0.07 0.11 0.07
1 year 33 to 329.999 mV 0.14 0.10 0.14 0.10
330 mV to 1020 V 0.12 0.08 0.12 0.08
Current Range [2]
Voltage Range
0.33 A to
0.8999 A
0.9 A to
2.1999 A
2.2 A to
4.4999 A
4.5 A to
20.5 A
Absolute Uncertainty, tcal ± 5 °C, ± (% of watts output) [1]
90 days 33 to 329.999 mV 0.12 0.10 0.12 0.10
330 mV to 1020 V 0.10 0.08 0.11 0.09
1 year 33 to 329.999 mV 0.13 0.11 0.13 0.11
330 mV to 1020 V 0.11 0.09 0.12 0.10
[1] To determine ac power uncertainty with more precision, see the individual “DC Voltage Specifications” and “DC
Current Specifications” and “Calculating Power Uncertainty.”
[2] Add 0.02% unless a settling time of 30 seconds is allowed for output currents > 10 A or for currents on the highest
two