EasyManua.ls Logo

Fluke 5520A - Leveled Sine Wave Harmonics Verification

Fluke 5520A
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5520A
Service Manual
6-42
6-59. Leveled Sine Wave Harmonics Verification
This procedure uses the following equipment:
Hewlett-Packard 8590A Spectrum Analyzer
BNC(f) to Type N(m) adapter
BNC cable supplied with the SC600
Refer to Figure 6-9 for proper setup connections.
HP 8590A
SC600
Cable
BNC(F)
to Type N (M)
Adapter
5520A-SC600
HI
LO
TRIG
GUARD
TC
20A
1000V
RMS
MAX
20V
RMS
MAX
20V
PK
MAX
1V PK
MAX
SHELLS
NOT
GROUNDED
NORMAL AUX
SCOPE
OUT
V, , ,
RTD
A, -SENSE,
AUX V
150V
PK
MAX
20V
RMS
MAX
20V PK MAX
20V PK MAX
5520A
CALIBRATOR
yg059f.eps
Figure 6-9. Leveled Sine Wave Harmonics Verification Setup
Set the Calibrator Mainframe to SCOPE mode, with the Levsine menu on the display.
Then follow these steps to verify the leveled sine wave harmonics.
1. Using the BNC cable and BNC(f) to Type N(m) adapter, connect the SCOPE
connector on the Calibrator Mainframe to the HP 8590A.
2. Program the Calibrator Mainframe to 5.5 V p-p at each frequency listed in Table 6-
31. Press O on the Calibrator Mainframe to activate the output.
3. Set HP 8590A start frequency to the Calibrator Mainframe output frequency. Set HP
8590A stop frequency to 10 times the Calibrator Mainframe output frequency. Set the
HP 8590A reference level at +19 dBm.
4. Record the harmonic level reading for each frequency and harmonic listed in Table 6-
31. For harmonics 3, 4, and 5, record the highest harmonic level of the three
measured. Harmonics should be below the levels listed in the tolerance column of
Table 6-31.

Table of Contents

Other manuals for Fluke 5520A

Related product manuals