EasyManua.ls Logo

Fluke 5520A - 6-57. Leveled Sine Wave Amplitude Verification

Fluke 5520A
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SC600 Option
Verification
6
6-39
Table 6-27. Edge Aberrations
Time from 50% of Rising Edge Typical Edge Aberrations
0 - 2 ns < 32 mV (3.2%)
2 - 5 ns < 22 mV (2.2%)
5 - 15 ns < 12 mV (1.2%)
> 15 ns < 7 mV (0.7%)
6-56. Tunnel Diode Pulser Drive Amplitude Verification
This procedure uses the following equipment:
Hewlett-Packard 3458A Digital Multimeter
BNC(f) to Double Banana adapter
BNC cable supplied with the SC600
Set the Calibrator Mainframe in Scope Cal mode, Edge. Proceed with the following
steps:
1. Connect the Calibrator Mainframe’s SCOPE connector to the HP 3458A input, using
the BNC cable and the BNC(f) to Double Banana adapter. Refer to Figure 6-2 for the
proper setup connections.
2. Activate the TD Pulser output by pushing the TDPULSE blue softkey. The output
should now be at 80 V peak-to-peak, 100 kHz, STANDBY.
3. Set the HP 3458A to DCV, NPLC = .001, LEVEL 1, TRIG LEVEL, and the DELAY
to .00012 for measuring the topline and DELAY to .00007 for measuring the
baseline. Manually range lock the HP 3458A to the 100 V dc range.
4. Change the Calibrator Mainframe output frequency to 10 kHz. Push the operate key,
and use the HP 3458A to measure the topline and baseline.
5. The peak-to-peak value is the difference between the topline and baseline. Record
these values in Table 6-28, and compare against the listed tolerance.
Table 6-28. Tunnel Diode Pulser Amplitude Verification
Calibrator
Mainframe
Edge Output
HP 3458A
Range
Topline
Reading
Baseline
Reading
Peak-to-Peak
Tolerance
(±V)
80 V, 10 kHz 100 V dc 1.6
6-57. Leveled Sine Wave Amplitude Verification
This procedure uses the following equipment:
5790A AC Measurement Standard
BNC(f) to Double Banana Plug adapter
50 feedthrough termination
BNC cable supplied with the SC600
Refer to Figure 6-17 for the proper setup connections.

Table of Contents

Other manuals for Fluke 5520A

Related product manuals