1-14. AC Current (Sine Wave) Specifications
LCOMP off
Max
Inductive
Load
Range Frequency
Absolute Uncertainty,
tcal ± 5 °C
± (% of output + µA)
Compliance
adder
± (µA/V)
Max
Distortion &
Noise 10 Hz
to 100 kHz
BW
± (% output
+
90 days 1 year floor) µH
29.00 µA to 10 Hz to 20 Hz 0.16 + 0.1 0.2 + 0.1 0.05 0.15 + 0.5 µA 200
329.99 µA 20 Hz to 45 Hz 0.12 + 0.1 0.15 + 0.1 0.05 0.1 + 0.5 µA
45 Hz to 1 kHz 0.1 + 0.1 0.125 + 0.1 0.05 0.05 + 0.5 µA
1 kHz to 5 kHz 0.25 + 0.15 0.3 + 0.15 1.5 0.5 + 0.5 µA
5 kHz to 10 kHz 0.6 + 0.2 0.8 + 0.2 1.5 1.0 + 0.5 µA
10 kHz to 30
kHz
1.2 + 0.4 1.6 + 0.4 10 1.2 + 0.5 µA
0.33 mA to 10 Hz to 20 Hz 0.16 + 0.15 0.2 + 0.15 0.05 0.15 + 1.5 µA 200
3.2999 mA 20 Hz to 45 Hz 0.1 + 0.15 0.125 + 0.15 0.05 0.06 + 1.5 µA
45 Hz to 1 kHz 0.08 + 0.15 0.1+ 0.15 0.05 0.02 + 1.5 µA
1 kHz to 5 kHz 0.16 + 0.2 0.2 + 0.2 1.5 0.5 + 1.5 µA
5 kHz to 10 kHz 0.4 + 0.3 0.5 + 0.3 1.5 1.0 + 1.5 µA
10 kHz to 30
kHz
0.8 + 0.6 1.0 + 0.6 10 1.2 + 0.5 µA
3.3 mA to 10 Hz to 20 Hz 0.15 + 2 0.18 + 2 0.05 0.15 + 5 µA50
32.999 mA 20 Hz to 45 Hz 0.075 + 2 0.09 + 2 0.05 0.05 + 5 µA
45 Hz to 1 kHz 0.035 + 2 0.04 + 2 0.05 0.07 + 5 µA
1 kHz to 5 kHz 0.065 + 2 0.08 + 2 1.5 0.3 + 5 µA
5 kHz to 10 kHz 0.16 + 3 0.2 + 3 1.5 0.7 + 5 µA
10 kHz to 30
kHz
0.32 + 4 0.4 + 4 10 1.0 + 0.5 µA
33 mA to 10 Hz to 20 Hz 0.15 +20 0.18 + 20 0.05 0.15 + 50 µA50
329.99 mA 20 Hz to 45 Hz 0.075 + 20 0.09 + 20 0.05 0.05 + 50 µA
45 Hz to 1 kHz 0.035 + 20 0.04 + 20 0.05 0.02 + 50 µA
1 kHz to 5 kHz 0.08 + 50 0.10 + 50 1.5 0.03 + 50 µA
5 kHz to 10 kHz 0.16 + 100 0.2 + 100 1.5 0.1 + 50 µA
10 kHz to 30
kHz
0.32 + 200 0.4 + 200 10 0.6 + 50 µA
0.33 A to 10 Hz to 45 Hz 0.15 + 100 0.18 + 100 0.2 + 500 µA2.5
1.09999 A 45 Hz to 1 kHz 0.036 + 100 0.05 + 100 0.07 + 500
µ
A
1 kHz to 5 kHz 0.5 + 1000 0.6 + 1000 [3] 1 + 500 µA
5 kHz to 10 kHz 2.0 + 5000 2.5 + 5000 [4] 2 + 500 µA
1.1 A to 10 Hz to 45 Hz 0.15 + 100 0.18 + 100 0.2 + 500 µA2.5
2.99999 A 45 Hz to 1 kHz 0.05 + 100 0.06 + 100 0.07 + 500
µ
A
1 kHz to 5 kHz 0.5 + 1000 0.6 + 1000 [3] 1 + 500 µA
5 kHz to 10 kHz 2.0 + 5000 2.5 + 5000 [4] 2 + 500 µA
3 A to 45 Hz to 100 Hz 0.05 + 2000 0.06 + 2000 0.2 + 3 mA 1
10.9999 A
100 kHz to 1
kHz
0.08 + 2000 0.10 + 2000 0.1 + 3 mA
1 kHz to 5 kHz 2.5 + 2000 3.0 + 2000 0.8 + 3 mA
11A to 45 Hz to 100 Hz 0.1 + 5000 0.12 + 5000 0.2 + 3 mA 1
20.5 A 100 Hz to 1 kHz 0.13 + 5000 0.15 + 5000 0.1 + 3 mA
[2] 1 kHz to 5 kHz 2.5 + 5000 3.0 + 5000 0.8 + 3 mA
[1] Max Distortion for 100 kHz to 200 kHz. For 200 kHz to 500 kHz, the maximum distortion is 0.9% of
output + floor as shown.
[2] Duty Cycle: Currents < 11 A may be provided continuously. For currents > 11 A, see Figure 1-4. The
current ma
y
be provided 60-T-I minutes an
y
60 minute period where T is the temperature in °C
(
room
temperature is about 23°C) and I is the output current in Amps. For example, 17 A, at 23°C could be
provided for 60-17-23 = 20 minutes each hour.
[3] For compliance voltages greater than 1 V, add 1 mA/V to the floor specification from 1 kHz to 5 kHz.
[4] For compliance voltages greater than 1 V, add 5 mA/V to the floor specification from 5 kHz to 10 kHz.