Introduction and Specifications
Specifications
1
1-13
AC Voltage (Sine Wave) Specifications (cont)
AUX (Auxiliary Output) [dual output mode only] [1]
Range Frequency
Absolute Uncertainty,
tcal ± 5 °C
± (% of output + µV)
Resolution
Max
Burden
Max Distortion
and Noise
10 Hz to
100 kHz
Bandwidth
90 days 1 year
± (% output +
floor)
10 mV to 10 Hz to 20 Hz 0.15 + 370 0.2 + 370 1 µV5 mA
0.2 + 200 µV
329.999 mV 20 Hz to 45 Hz 0.08 + 370 0.1 + 370 0.06 + 200 µV
45 Hz to 1 kHz 0.08 + 370 0.1 + 370 0.08 + 200 µV
1 kHz to 5 kHz 0.15 + 450 0.2 + 450 0.3 + 200 µV
5 kHz to 10 kHz 0.3 + 450 0.4 + 450 0.6 + 200 µV
10 Hz to 30 kHz 4.0 + 900 5.0 + 900 1 + 200 µV
0.33 V to 10 Hz to 20 Hz 0.15 + 450 0.2 + 450 10 µV5 mA
0.2 + 200 µV
3.29999 V 20 Hz to 45 Hz 0.08 + 450 0.1 + 450 0.06 + 200 µV
45 Hz to 1 kHz 0.07 + 450 0.09 +
450
0.08 + 200 µV
1 kHz to 5 kHz 0.15 + 1400 0.2 +
1400
0.3 + 200 µV
5 kHz to 10 kHz 0.3 + 1400 0.4 +
1400
0.6 + 200 µV
10 kHz to 30 kHz 4.0 + 2800 5.0 +
2800
1 + 200 µV
3.3 V to 5 V 10 Hz to 20 Hz 0.15 + 450 0.2 + 450 100 µV5 mA
0.2 + 200 µV
20 Hz to 45 Hz 0.08 + 450 0.1 + 450 0.06 + 200 µV
45 Hz to 1 kHz 0.07 + 450 0.09 +
450
0.08 + 200 µV
1 kHz to 5 kHz 0.15 + 1400 0.2 +
1400
0.3 + 200 µV
5 kHz to 10 kHz 0.3 +1400 0.4 +
1400
0.6 + 200 µV
[1] There are two channels of voltage output. The maximum frequency of the dual output is 30 kHz.
Note
Remote sensing is not provided. Output resistance is < 5 m for outputs 0.33 V. The AUX output
resistance is < 1. The maximum load capacitance is 500 pF, subject to the maximum burden
current
limits.

Table of Contents