Calibration and Verification
Performance Verification Tests
3
3-35
3-19. Verifying DC Volts (NORMAL Output)
Verify that the 5520A performance is within the limits in Table 3-19, using the same
equipment and techniques specified previously for calibration.
Table 3-19. Verification Tests for DC Voltage (NORMAL Output)
Range Output Lower Limit Upper Limit
329.9999 mV 0.0000 mV -0.0010 mV 0.0010 mV
329.9999 mV 329.0000 mV 328.9941 mV 329.0059 mV
329.9999 mV -329.0000 mV -329.0059 mV -328.9941 mV
3.299999 V 0.000000 V -0.000002 V 0.000002 V
3.299999 V 1.000000 V 0.999989 V 1.000011 V
3.299999 V -1.000000 V -1.000011 V -0.999989 V
3.299999 V 3.290000 V 3.289968 V 3.290032 V
3.299999 V -3.290000 V -3.290032 V -3.289968 V
32.99999 V 0.00000 V -0.00002 V 0.00002 V
32.99999 V 10.00000 V 9.99988 V 10.00012 V
32.99999 V -10.00000 V -10.00012 V -9.99989 V
32.99999 V 32.90000 V 32.89965 V 32.90035 V
32.99999 V -32.90000 V -32.90035V -32.89965 V
329.9999 V 50.0000 V 49.9991 V 50.0009 V
329.9999 V 329.0000 V 328.9949 V 329.0051 V
329.9999 V -50.0000 V -50.0009 V -49.9991 V
329.9999 V -329.0000 V -329.0051 V -328.9949 V
1000.000 V 334.000 V 333.993 V 334.007 V
1000.000 V 900.000 V 899.985 V 900.015 V
1000.000 V 1020.000 V 1019.983 V 1020.017 V
1000.000 V -334.000 V -334.007 V -333.993 V
1000.000 V -900.000 V -900.015 V -899.985 V
1000.000 V -1020.000 V -1020.017 V -1019.983 V