EasyManua.ls Logo

Fluke 5520A - 6-134. Verification at 1 MΩ; 6-135. Verification at 50 Ω

Fluke 5520A
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SC300 Option
Verification
6
6-105
SC300
Cable
5520A-SC300
50
Feed Through
Termination
BNC (F) to
Double Banana
Adapter
HI
LO
TRIG
GUARD
TC
20A
1000V
RMS
MAX
20V
RMS
MAX
20V
PK
MAX
1V PK
MAX
SHELLS
NOT
GROUNDED
NORMAL AUX
SCOPE
OUT
V, , ,
RTD
A, -SENSE,
AUX V
150V
PK
MAX
20V
RMS
MAX
20V PK MAX
20V PK MAX
5520A
CALIBRATOR
yg126f.eps
Figure 6-28. Wave Generator Verification Setup
Set the Calibrator Mainframe to SCOPE mode, with the Wavegen menu on the display.
Press O on the Calibrator Mainframe to activate the output. Set the offset to 0 mV,
and the frequency to 1 kHz. Then follow these steps to verify the wave generator
function.
6-134. Verification at 1 M
1. Set the Calibrator Mainframe impedance to 1 M (The blue softkey under SCOPE Z
toggles the impedance between 50 and 1 M).
2. Connect the BNC cable to the Calibrator Mainframe’s SCOPE connector. Connect
the other end of the BNC cable to the 5790A INPUT 2 using the BNC(f) to Double
Banana adapter.
3. Set the 5790A to AUTORANGE, digital filter mode to FAST, restart fine, and Hi
Res on.
4. Program the Calibrator Mainframe to output the wave type and voltage listed in
Table 6-58.
5. Allow the 5790A reading to stabilize, then record the 5790A rms reading for each
wave type and voltage in Table 6-58.
6. Multiply the rms reading by the conversion factor listed to convert it to the peak-to-
peak value. Compare result to the tolerance column.
6-135. Verification at 50
1. Set the Calibrator Mainframe impedance to 50 (The blue softkey under SCOPE Z
toggles the impedance between 50 and 1 M).
2. Connect the BNC cable to the Calibrator Mainframe’s SCOPE connector. Connect
the other end of the BNC cable to the 50 feedthrough termination then to the
5790A INPUT 2 using the BNC(f) to Double Banana adapter.
3. Set the 5790A to AUTORANGE, digital filter mode to FAST, restart fine, and Hi
Res on.