1-20. Power and Dual Output Limit Specifications
Frequency
Voltages
(NORMAL)
Currents
Voltages
(AUX)
Power
Factor
(PF)
dc 0 to ± 1020 V 0 to ± 20.5 A 0 to ± 7 V
10 Hz to 45 Hz 33 mV to 32.9999 V 3.3 mA to 2.99999 A 10 mV to 5 V 0 to 1
45 Hz to 65 Hz 33 mV to 1000 V 3.3 mA to 20.5 A 10 mV to 5 V 0 to 1
65 Hz to 500 Hz 330 mV to 1000 V 33 mA to 2.99999 A 100 mV to 5 V 0 to 1
65 Hz to 500 Hz 3.3 V to 1000 V 33 mA to 20.5 A 100 mV to 5 V 0 to 1
500 Hz to 1 kHz 330 mV to 1000 V 33 mA to 20.5 A 100 mV to 5 V 1
1 kHz to 5 kHz 3.3 V to 1000 V [1] 33 mA to 2.99999 A 100 mV to 5 V [1] 1
5 kHz to 10 kHz 3.3 V to 1000 V [2] 33 mA to 329.99 mA 1 V to 5 V [2] 1
[1] In dual voltage output mode, voltage is limited to 3.3 V to 500 V in the NORMAL output.
[2] In dual voltage output mode, voltage is limited to 3.3 V to 250 V in the NORMAL output.
Notes
The range of voltages and currents shown in “DC Voltage Specifications,” “DC Current Specifications,” “AC
Voltage (Sine Wave) Specifications,” and “AC Current (Sine Wave) Specifications” are available in the
power and dual output modes (except minimum current for ac power is 0.33 mA). However, only those
limits shown in this table are specified. See “Calculating Power Uncertainty” to determine the uncertainty at
these points.
The phase adjustment range for dual ac outputs is 0° to ± 179.99°. The phase resolution for dual ac
outputs is 0.01°.