Fluke 5520A - 3-14. Capacitance Calibration

Fluke 5520A
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5520A
Service Manual
3-24
HI
LO
TRIG
GUARD
TC
20A
1000V
RMS
MAX
20V
RMS
MAX
20V
PK
MAX
1V PK
MAX
SHELLS
NOT
GROUNDED
NORMAL AUX
SCOPE
OUT
V, , ,
RTD
A, -SENSE,
AUX V
150V
PK
MAX
20V
RMS
MAX
20V PK MAX
20V PK MAX
5520A
CALIBRATOR
UUT
HP3458A
Set the HP3458A
to external guard
yg113f.eps
Figure 3-12. Two-Wire Resistance Connection
HP3458A
P1 P2
Set the HP3458A
to external guard
yg114f.eps
Figure 3-13. Scaling the DMM to a Guildline 9334
3-14. Capacitance Calibration
The equipment listed in Table 3-16 is required for calibration of the resistance function.
(The equipment is also listed in the consolidated table, Table 3-1.)
Table 3-16. Test Equipment Required for Calibrating Capacitance
Quan. Manufacturer Model Equipment
1 Fluke PM 9540/BAN Cable Set
1 Fluke PM 6304C LCR Meter