5520A
Service Manual
6-8
6-6. Leveled Sine Wave Specifications
Table 6-3. Leveled Sine Wave Specifications
Leveled Sine Wave
Frequency Range
Characteristics
into 50Ω
50 kHz
(reference)
50 kHz to
100 MHz
100 MHz to
300 MHz
300 MHz to
600 MHz
Amplitude Characteristics (for measuring oscilloscope bandwidth)
Range (p-p) 5 mV to 5.5 V
Resolution < 100 mV:3 digits
≥ 100 mV:4 digits
Adjustment Range continuously adjustable
1-Year Absolute
Uncertainty,
tcal ± 5 °C
± (2% of
output
+ 300 µV)
± (3.5% of
output
+ 300 µV)
± (4% of output
+ 300 µV)
± (6% of output
+ 300 µV)
Flatness (relative to
50 kHz)
not applicable ± (1.5% of
output
+ 100 µV)
± (2% of
output
+ 100 µV)
± (4% of output
+ 100 µV)
Short-Term Amplitude
Stability
≤ 1% [1]
Frequency Characteristics
Resolution 10 kHz
1-Year Absolute
Uncertainty,
tcal ± 5 °C
± 2.5 ppm
Distortion Characteristics
2nd Harmonic ≤ -33 dBc
3rd and Higher
Harmonics
≤ -38 dBc
[1] Within one hour after reference amplitude setting, provided temperature varies no more than ± 5°C.