6-127. Leveled Sine Wave Flatness Verification
Leveled Sine Wave flatness verification is divided into two frequency bands: 50 kHz to
10 MHz (low frequency) and > 10 MHz to 300 MHz (high frequency). The equipment
setups are different for each band. Leveled Sine Wave flatness is measured relative to 50
kHz. This is determined directly in the low frequency band. The high frequency band
requires a “transfer” measurement be made at 10 MHz to calculate a flatness relative to
50 kHz.
6-128. Equipment Setup for Low Frequency Flatness
All low frequency flatness procedures use the following equipment:
5790A/03 AC Measurement Standard with Wideband option
BNC(f) to Type N(m) adapter
BNC cable supplied with the SC300
Connect the Calibrator Mainframe SCOPE connector to the 5790A WIDEBAND input
with the BNC(f) to Type N(m) adapter as shown in Figure 6-25. Set the 5790A to
AUTORANGE, digital filter mode to FAST, restart fine, and Hi Res on.
POWER
I
O
5790A
AC MEASUREMENT
STANDARD
1000V RMS MAX 1000V RMS MAX
3V RMS MAX
INPUT 1 INPUT 2
SHUNT
HI
LO
GUARDGROUND
SHELL FLOATING
10V PEAK
MAX
WIDEBAND
7V RMS MAX
MAX
10V PK
SHELL FLOATING
INPUT1
SHUNT
INPUT1 INPUT1 INPUT1
2.2 mV 22 mV 220 mV
7 mV 70 mV 700 mV
2.2 V 22 V 220 mV
7V 70V 700V
1kV 2.2 mV
AUTO
MAN
VIEW
REF
UTIL
MENUS
SPEC
CLEAR
DELETE
ENTER
01 2 3
4
5
6
7
89
+/-
.
POWER
I
O
0•
123
456
789
ENTER
M
k
m
VHz
FIELD
EDIT
/
+
F
OPR EARTH SCOPE BOOST MENU
PREV
SHIFT
RESET
CE
SETUP
REF
NEW
TC
MEAS
¡F
µ
n
p
W
dBm sec
¡CA
MULT
x
DIV
÷
OUT
TRIG
5520A
CALIBRATOR
STBY
HI
LO
TRIG
GUARD
TC
20A
1000V
RMS
MAX
20V
RMS
MAX
20V
PK
MAX
1V PK
MAX
SHELLS
NOT
GROUNDED
NORMAL AUX
SCOPE
OUT
V, , ,
RTD
A, -SENSE,
AUX V
150V
PK
MAX
20V
RMS
MAX
20V PK MAX
20V PK MAX
yg034f.eps
Figure 6-25. Connecting the Calibrator Mainframe to the 5790A AC Measurement Standard
6-129. Equipment Setup for High Frequency Flatness
All high frequency flatness procedures use the following equipment:
Hewlett-Packard 437B Power Meter
Hewlett-Packard 8482A and 8481D Power Sensors
BNC(f) to Type N(f) adapter
BNC cable supplied with the Calibrator Mainframe
Note
When high frequencies at voltages below 63 mV p-p are verified, use the
8481D Power Sensor. Otherwise, use the 8482A Power Sensor.