5520A
Service Manual
6-100
Connect the HP 437B Power Meter to either the 8482A or the 8481D Power Sensor as
shown in Figure 6-26. For more information on connecting the two instruments, see the
power meter and power sensor operators manuals.
Connect the power meter/power sensor combination to the SCOPE connector on the
Calibrator Mainframe, as shown in Figure 6-27.
The Hewlett-Packard 437B Power Meter must be configured by setting the parameters
listed below. Zero and self-calibrate the power meter with the power sensor being used.
Refer to the Hewlett-Packard 437B operators manual for details.
• PRESET
• RESOLN 3
• AUTO FILTER
• WATTS
• SENSOR TABLE 0 (default)
OM035f.eps
Figure 6-26. Connecting the HP 437B Power Meter to the HP 8482A or 8481D Power Sensor