5520A
Service Manual
1-28
1-29. AC Voltage, Square Wave Characteristics
Risetime
@ 1 kHz
Typical
Settling Time
@ 1 kHz
Typical
Overshoot
@ 1 kHz
Typical
Duty Cycle
Range
Duty Cycle Uncertainty
< 1 µs < 10 µs to 1%
of final value
< 2% 1% to 99%,
< 3.3 V p-p,
0.01 Hz to
100 kHz
± (0.02% of period + 100 ns), 50% duty cycle
± (0.05% of period + 100 ns), other duty cycles
from 10% to 90%
1-30. AC Voltage, Triangle Wave Characteristics (typical)
Linearity to 1 kHz Aberrations
0.3% of p-p value, from 10% to 90% point < 1% of p-p value, with amplitude >50% of range
1-31. AC Current (Sine Wave) Extended Bandwidth Specifications
Range Frequency
1-Year Absolute Uncertainty
tcal ± 5 °C
± (% of output + % of range) [2]
Max
Current
Resolution
All current ranges, < 330 mA 0.01 Hz to 10 Hz 5.0 + 0.5 2 digits