3-23. Verifying AC Voltage (NORMAL Output)
Verify that the 5520A performance is within the limits in Table 3-24, using the same
equipment and techniques specified previously for calibration.
Table 3-24. Verification Tests for AC Voltage (NORMAL Output)
Range Output Frequency Lower Limit Upper Limit
32.999 mV 3.000 mV 45 Hz 2.994 mV 3.006 mV
32.999 mV 3.000 mV 10 kHz 2.994 mV 3.006 mV
32.999 mV 30.000 mV 9.5 Hz 28.335 mV 31.665 mV
32.999 mV 30.000 mV 10 Hz 29.976 mV 30.024 mV
32.999 mV 30.000 mV 45 Hz 29.990 mV 30.010 mV
32.999 mV 30.000 mV 1 kHz 29.990 mV 30.010 mV
32.999 mV 30.000 mV 10 kHz 29.990 mV 30.010 mV
32.999 mV 30.000 mV 20 kHz 29.989 mV 30.011 mV
32.999 mV 30.000 mV 50 kHz 29.970 mV 30.030 mV
32.999 mV 30.000 mV 100 kHz 29.898 mV 30.102 mV
32.999 mV 30.000 mV 450 kHz 29.770 mV 30.230 mV
329.999 mV 33.000 mV 45 Hz 32.987 mV 33.013 mV
329.999 mV 33.000 mV 10 kHz 32.987 mV 33.013 mV
329.999 mV 300.000 mV 9.5 Hz 283.350 mV 316.650 mV
329.999 mV 300.000 mV 10 Hz 299.917 mV 300.083 mV
329.999 mV 300.000 mV 45 Hz 299.950 mV 300.050 mV
329.999 mV 300.000 mV 1 kHz 299.950 mV 300.050 mV
329.999 mV 300.000 mV 10 kHz 299.950 mV 300.050 mV
329.999 mV 300.000 mV 20 kHz 299.947 mV 300.053 mV
329.999 mV 300.000 mV 50 kHz 299.902 mV 300.098 mV
329.999 mV 300.000 mV 100 kHz 299.788 mV 300.212 mV
329.999 mV 300.000 mV 500 kHz 299.450 mV 300.550 mV
3.29999 V 0.33000 V 45 Hz 0.32989 V 0.33011 V
3.29999 V 0.33000 V 10 kHz 0.32989 V 0.33011 V
3.29999 V 3.00000 V 9.5 Hz 2.83350 V 3.16650 V
3.29999 V 3.00000 V 10 Hz 2.99920 V 3.00080 V
3.29999 V 3.00000 V 45 Hz 2.99952 V 3.00048 V
3.29999 V 3.00000 V 1 kHz 2.99952 V 3.00048 V
3.29999 V 3.00000 V 10 kHz 2.99952 V 3.00048 V
3.29999 V 3.00000 V 20 kHz 2.99946 V 3.00054 V