Calibration and Verification
Performance Verification Tests
3
3-39
Table 3-23. Verification Tests for Resistance (cont)
Range Output Lower Limit Upper Limit
3.299999 MΩ 1.190000 MΩ 1.189922 MΩ 1.190078 MΩ
3.299999 MΩ 1.900000 MΩ 1.899894 MΩ 1.900106 MΩ
3.299999 MΩ 3.000000 MΩ 2.999850 MΩ 3.000150 MΩ
10.99999 MΩ 3.30000 MΩ 3.29959 MΩ 3.30041 MΩ
10.99999 MΩ 10.90000 MΩ 10.89875 MΩ 10.90125 MΩ
32.99999 MΩ 11.90000 MΩ 11.89512 MΩ 11.90488 MΩ
32.99999 MΩ 19.00000 MΩ 18.99370 MΩ 19.00630 MΩ
32.99999 MΩ 30.00000 MΩ 29.99150 MΩ 30.00850 MΩ
109.9999 MΩ 33.0000 MΩ 32.9838 MΩ 33.0162 MΩ
109.9999 MΩ 109.0000 MΩ 108.9534 MΩ 109.0466 MΩ
329.9999 MΩ 119.0000 MΩ 118.6025 MΩ 119.3975 MΩ
329.9999 MΩ 290.0000 MΩ 289.1750 MΩ 290.8250 MΩ
1100.000 MΩ 400.000 MΩ 394.700 MΩ 405.300 MΩ
1100.000 MΩ 640.000 MΩ 631.820 MΩ 648.180 MΩ
1100.000 MΩ 1090.000 MΩ 1076.420 MΩ 1103.580 MΩ