SC600 Option
SC600 Specifications
6
6-11
6-12. Trigger Signal Specifications (Edge Function)
Table 6-9. Trigger Signal Specifications (Edge Function)
Edge Signal
Frequency
Division
Ratio
Typical Amplitude
into 50 (p-p)
Typical Rise Time Typical Lead Time
1 kHz to 10 MHz off/1 1 V 2 ns 40 ns
6-13. Trigger Signal Specifications (Square Wave Voltage Function)
Table 6-10. Trigger Signal Specifications (Square Wave Voltage Function)
Edge Signal
Frequency
Division
Ratio
Typical Amplitude
into 50 (p-p)
Typical Rise Time Typical Lead Time
10 Hz to 10 kHz off/1 1 V 2 ns 1 µs
6-14. Trigger Signal Specifications
Table 6-11. TV Trigger Signal Specifications
Trigger Signal Type Parameters
Field Formats Selectable NTSC, SECAM, PAL, PAL-M
Polarity Selectable inverted or uninverted video
Amplitude into 50 (p-p) Adjustable 0 to 1.5 V p-p into 50 ohm load, (±7% accuracy)
Line Marker Selectable Line Video Marker
6-15. Oscilloscope Input Resistance Measurement Specifications
Table 6-12. Oscilloscope Input Resistance Measurement Specifications
Scope input selected 50 1 M
Measurement Range 40 to 60 500 k to 1.5 M
Uncertainty 0.1 % 0.1 %
6-16. Oscilloscope Input Capacitance Measurement Specifications
Table 6-13. Oscilloscope Input Capacitance Measurement Specifications
Scope input selected 1 M
Measurement Range 5 pF to 50 pF
Uncertainty ± (5% of input + 0.5 pF) [1]
[1] Measurement made within 30 minutes of capacitance zero reference. SC600 option must be
selected for at least five minutes prior to any capacitance measurement, including the zero process.