Fluke 5520A - 6-118. AC Voltage Frequency Verification

Fluke 5520A
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SC300 Option
Verification
6
6-89
Table 6-46. AC Voltage Verification at 50 (cont.)
Nominal Value
(p-p)
Frequency
Measured Value
(p-p)
Deviation
(mV)
1-Year Spec.
(mV)
500.0 mV 10 kHz 1.35
1.0 V 100 Hz 2.60
1.0 V 1 kHz 2.60
1.0 V 10 kHz 2.60
2.0 V 10 Hz 5.10
2.0 V 100 Hz 5.10
2.0 V 1 kHz 5.10
2.0 V 5 kHz 5.10
2.0 V 10 kHz 5.10
6-118. AC Voltage Frequency Verification
Refer to Figure 6-21 for the proper setup connections.
This procedure uses the following equipment:
PM 6680 Frequency Counter with an TCXO timebase (Option PM 9678 or
equivalent)
BNC cable supplied with the SC300
At 50 MHZ
SC300 Cable
PM 6680A
5520A-SC300
HI
LO
TRIG
GUARD
TC
20A
1000V
RMS
MAX
20V
RMS
MAX
20V
PK
MAX
1V PK
MAX
SHELLS
NOT
GROUNDED
NORMAL AUX
SCOPE
OUT
V, , ,
RTD
A, -SENSE,
AUX V
150V
PK
MAX
20V
RMS
MAX
20V PK MAX
20V PK MAX
5520A
CALIBRATOR
yg123f.eps
Figure 6-21. Frequency Verification Setup
Set the Calibrator Mainframe to SCOPE mode, with the Volt menu on the display. Press
O on the Calibrator Mainframe to activate the output. Then follow these steps to
verify ac voltage frequency: