EasyManua.ls Logo

Fluke 5520A - 1-13. AC Voltage (Sine Wave) Specifications

Fluke 5520A
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5520A
Service Manual
1-12
1-13. AC Voltage (Sine Wave) Specifications
NORMAL (Normal Output)
Range Frequency
Absolute Uncertainty,
tcal ± 5 °C
± (ppm of output + µV)
Resolution
Max
Burden
Max Distortion
and Noise
10 Hz to 5 MHz
Bandwidth
90 days 1 year
± (% output +
floor)
1.0 mV to 10 Hz to 45 Hz 600 + 6 800 + 61 µV 50
0.15 + 90 µV
32.999 mV 45 Hz to 10 kHz 120 + 6 150 + 6 0.035 + 90 µV
10 kHz to 20 kHz 160 + 6 200 + 6 0.06 + 90 µV
20 kHz to 50 kHz 800 + 6 1000 + 6 0.15 + 90 µV
50 kHz to 100 kHz 3000 + 12 3500 + 12 0.25 + 90 µV
100 kHz to
500 kHz
6000 + 50 8000 + 50 0.3 + 90 µV [1]
33 mV to 10 Hz to 45 Hz 250 + 8 300 + 81 µV 50
0.15 + 90 µV
329.999 mV 45 Hz to 10 kHz 140 + 8 145 + 8 0.035 + 90 µV
10 kHz to 20 kHz 150 + 8 160 + 8 0.06 + 90 µV
20 kHz to 50 kHz 300 + 8 350 + 8 0.15 + 90 µV
50 kHz to 100 kHz 600 + 32 800 + 32 0.20 + 90 µV
100 kHz to
500 kHz
1600 + 70 2000 + 70 0.20 + 90 µV [1]
0.33 V to 10 Hz to 45 Hz 250 + 50 300 + 50 10 µV 10 mA
0.15 + 200 µV
3.29999 V 45 Hz to 10 kHz 140 + 60 150 + 60 0.035 + 200 µV
10 kHz to 20 kHz 160 + 60 190 + 60 0.06 + 200 µV
20 kHz to 50 kHz 250 + 50 300 + 50 0.15 + 200 µV
50 kHz to 100 kHz 550 + 125 700 + 125 0.20 + 200 µV
100 kHz to
500 kHz
2000 + 600 2400 + 600 0.20 + 200 µV [1]
3.3 V to 10 Hz to 45 Hz 250 + 650 300 + 650 100 µV 10 mA
0.15 + 2 mV
32.9999 V 45 Hz to 10 kHz 125 + 600 150 + 600 0.035 + 2 mV
10 kHz to 20 kHz 220 + 600 240 + 600 0.08 + 2 mV
20 kHz to 50 kHz 300 + 600 350 + 600 0.2 + 2 mV
50 kHz to 100 kHz 750 + 1600 900 + 1600 0.5 + 2 mV
33 V to 45 Hz to 1 kHz 150 + 2000 190 + 2000 1 mV 5 mA,
0.15 + 10 mV
329.999 V 1 kHz to 10 kHz 160 + 6000 200 + 6000 except 0.05 + 10 mV
10 kHz to 20 kHz 220 + 6000 250 + 6000 20 mA for 0.6 + 10 mV
20 kHz to 50 kHz 240 + 6000 300 + 6000 45 Hz to 0.8 + 10 mV
50 kHz to 100 kHz 1600 +
50000
2000 +
50000
65 Hz 1.0 + 10 mV
330 V to
1020 V
45 Hz to 1 kHz 250 + 10000 300 +
10000
10 mV 2 mA,
except
0.15 + 30 mV
1 kHz to 5 kHz 200 + 10000
250 +
10000
6 mA for
45 Hz to
0.07 + 30 mV
5 kHz to 10 kHz 250 + 10000 300 +
10000
65 Hz 0.07 + 30 mV
[1] Max Distortion for 100 kHz to 200 kHz. For 200 kHz to 500 kHz, the maximum distortion is 0.9% of
output + floor as shown.
Note
Remote sensing is not provided. Output resistance is < 5 m for outputs 0.33 V. The AUX output
resistance is < 1. The maximum load capacitance is 500 pF, subject to the maximum burden current
limits.