EasyManua.ls Logo

Fluke 5520A - 6-104. DC Voltage Calibration

Fluke 5520A
254 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5520A
Service Manual
6-78
HP 3458A 5520A-SC300
SC300 Cable
(50 Feedthrough
Termination as
required by the
calibration procedure)
BNC(F) to
Double Banana
Adapter
HI
LO
TRIG
GUARD
TC
20A
1000V
RMS
MAX
20V
RMS
MAX
20V
PK
MAX
1V PK
MAX
SHELLS
NOT
GROUNDED
NORMAL AUX
SCOPE
OUT
V, , ,
RTD
A, -SENSE,
AUX V
150V
PK
MAX
20V
RMS
MAX
20V PK MAX
20V PK MAX
5520A
CALIBRATOR
yg122f.eps
Figure 6-19. Equipment Setup for SC300 Square Wave Measurements.
For all measurements, the HP 3458A is in DCV, manual ranging, with level triggering
enabled. A convenient method to make these measurements from the HP 3458A’s front
panel is to program these settings into several of the user defined keys on its front panel.
For example, to make topline measurements at 1 kHz, you would set the DMM to
“NPLC .01; LEVEL 1; DELAY .0002; TRIG LEVEL”. To find the average of multiple
readings, you can program one of the keys to “MATH OFF; MATH STAT” and then use
the “RMATH MEAN” function to recall the average or mean value. Refer to Figure 6-19
for the proper connections.
6-104. DC Voltage Calibration
This procedure uses the following equipment:
Hewlett-Packard 3458A Digital Multimeter
50 feedthrough termination (as required in the calibration procedure)
Shorted Dual Banana Connector
BNC(f) to Double Banana adapter
BNC cable supplied with the SC300
Note
Full calibration of the Voltage Function requires both dc and ac
calibration.
Refer to Figure 6-19 for the proper setup connections.
Set the Calibrator Mainframe in Scope Cal mode, DC Voltage section. Follow these
steps to calibrate DC Voltage:
1. Connect the Calibrator Mainframe’s SCOPE connector to the HP 3458A input, using
the BNC cable and the BNC(f) to Double Banana adapter.
2. Set the HP 3458A to DCV, Auto Range, NPLC = 10, FIXEDZ = on.
3. Press the GO ON blue softkey.

Table of Contents