EasyManua.ls Logo

Fluke 5520A - 1-15. Capacitance Specifications

Fluke 5520A
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5520A
Service Manual
1-16
1-15. Capacitance Specifications
Absolute Uncertainty, tcal
± 5 °C
± (% of output + floor)
Allowed Frequency or
Charge-Discharge Rate
Range
90 days 1 year
Resolution
Min and Max
to Meet
Specification
Typical
Max for
< 0.5%
Error
Typical
Max for
< 1%
Error
0.19 nF to
0.3999 nF
0.38 + 0.01 nF 0.5 + 0.01 nF 0.1 pF 10 Hz to 10 kHz 20 kHz 40 kHz
0.4 nF to
1.0999 nF
0.38 + 0.01 nF 0.5 + 0.01 nF 0.1 pF 10 Hz to 10 kHz 30 kHz 50 kHz
1.1 nF to
3.2999 nF
0.38 + 0.01 nF 0.5 + 0.01 nF 0.1 pF 10 Hz to 3 kHz 30 kHz 50 kHz
3.3 nF to
10.9999 nF
0.19 + 0.01 nF 0.25 + 0.01 nF 0.1 pF 10 Hz to 1 kHz 20 kHz 25 kHz
11 nF to
32.9999 nF
0.19 + 0.1 nF 0.25 + 0.1 nF 0.1 pF 10 Hz to 1 kHz 8 kHz 10 kHz
33 nF to
109.999 nF
0.19 + 0.1 nF 0.25 + 0.1 nF 1 pF 10 Hz to 1 kHz 4 kHz 6 kHz
110 nF to
329.999 nF
0.19 + 0.3 nF 0.25 + 0.3 nF 1 pF 10 Hz to 1 kHz 2.5 kHz 3.5 kHz
0.33 µF to
1.09999 µF
0.19 + 1 nF 0.25 + 1 nF 10 pF 10 Hz to 600 Hz 1.5 kHz 2 kHz
1.1 µF to
3.29999 µF
0.19 + 3 nF 0.25 + 3 nF 10 pF 10 Hz to 300 Hz 800 Hz 1 kHz
3.3 µF to
10.9999 µF
0.19 + 10 nF 0.25 + 10 nF 100 pF 10 Hz to 150 Hz 450 Hz 650 Hz
11 µF to
32.9999 µF
0.30 + 30 nF 0.40 + 30 nF 100 pF 10 Hz to 120 Hz 250 Hz 350 Hz
33 µF to
109.999 µF
0.34 + 100 nF 0.45 + 100 nF 1 nF 10 Hz to 80 Hz 150 Hz 200 Hz
110 µF to
329.999 µF
0.34 + 300 nF 0.45 + 300 nF 1 nF 0 to 50 Hz 80 Hz 120 Hz
0.33 µF to
1.09999mF
0.34 + 1 µF 0.45 + 1 µF 10 nF 0 to 20 Hz 45 Hz 65 Hz
1.1 mF to
3.2999 mF
0.34 + 3 µF 0.45 + 3 µF 10 nF 0 to 6 Hz 30 Hz 40 Hz
3.3 mF to
10.9999 mF
0.34 + 10 µF 0.45 + 10 µF 100 nF 0 to 2 Hz 15 Hz 20 Hz
11 mF to
32.9999 mF
0.7 + 30 µF 0.75 + 30 µF 100 nF 0 to 0.6 Hz 7.5 Hz 10 Hz
33 mF to
110 mF
1.0 + 100 µF1.1 + 100 µF 10 µF 0 to 0.2 Hz 3 Hz 5 Hz
Notes:
1. The output is continuously variable from 190 pF to 110 mF.
2. Specifications apply to both dc charge/discharge capacitance meters and ac RCL meters. The
maximum allowable peak voltage is 3 V. The maximum allowable peak current is 150 mA, with an rms
limitation of 30 mA below 1.1 µF and 100 mA for 1.1 µF and above.
3. The maximum lead resistance for no additional error in 2-wire COMP mode is 10 .

Table of Contents