3-24. Verifying AC Voltage (AUX Output)
Verify that the 5520A performance is within the limits in Table 3-25, using the same
equipment and techniques specified previously for calibration.
Table 3-25. Verification Tests for AC Voltage (AUX Output)
Range
Output, AUX
(Note)
Frequency Lower Limit Upper Limit
329.999 mV 10.000 mV 45 Hz 9.622 mV 10.378 mV
329.999 mV 10.000 mV 1 kHz 9.622 mV 10.378 mV
329.999 mV 10.000 mV 5 kHz 9.535 mV 10.465 mV
329.999 mV 10.000 mV 10 kHz 9.520 mV 10.480 mV
329.999 mV 10.000 mV 30 kHz 8.700 mV 11.300 mV
329.999 mV 300.000 mV 9.5 Hz 283.350 mV 316.650 mV
329.999 mV 300.000 mV 10 Hz 299.180 mV 300.820 mV
329.999 mV 300.000 mV 45 Hz 299.390 mV 300.610 mV
329.999 mV 300.000 mV 1 kHz 299.390 mV 300.610 mV
329.999 mV 300.000 mV 5 kHz 299.100 mV 300.900 mV
329.999 mV 300.000 mV 10 kHz 298.650 mV 301.350 mV
329.999 mV 300.000 mV 30 kHz 287.100 mV 312.900 mV
3.29999 V 3.00000 V 9.5 Hz 2.825 V 3.175 V
3.29999 V 3.00000 V 10 Hz 2.99505 V 3.00495 V
3.29999 V 3.00000 V 45 Hz 2.99745 V 3.00255 V
3.29999 V 3.00000 V 1 kHz 2.99745 V 3.00255 V
3.29999 V 3.00000 V 5 kHz 2.99410 V 3.00590 V
3.29999 V 3.00000 V 10 kHz 2.98960 V 3.01040 V
3.29999 V 3.00000 V 30 kHz 2.87720 V 3.12280 V
5.00000 V 5.00000 V 9.5 Hz 4.72500 V 5.27500 V
5.00000 V 5.00000 V 10 Hz 4.99205 V 5.00795 V
5.00000 V 5.00000 V 45 Hz 4.99605 V 5.00395 V
5.00000 V 5.00000 V 1 kHz 4.99605 V 5.00395 V
5.00000 V 5.00000 V 5 kHz 4.99110 V 5.00890 V
5.00000 V 5.00000 V 10 kHz 4.98360 V 5.01640 V
Note: set the NORMAL output to 300 mV.