Calibration and Verification
Performance Verification Tests
3
3-41
Table 3-24. Verification Tests for AC Voltage (NORMAL Output) (cont)
Range Output Frequency Lower Limit Upper Limit
3.29999 V 3.00000 V 50 kHz 2.99920 V 3.00080 V
3.29999 V 3.00000 V 100 kHz 2.99823 V 3.00178 V
3.29999 V 3.00000 V 450 kHz 2.99340 V 3.00660 V
3.29999 V 3.29000 V 2 MHz 0.07500 V (Note)
32.9999 V 3.3000 V 45 Hz 3.2990 V 3.3010 V
32.9999 V 3.3000 V 10 kHz 3.2990 V 3.3010 V
32.9999 V 30.0000 V 9.5 Hz 28.3350 V 31.6650 V
32.9999 V 30.0000 V 10 Hz 29.9919 V 30.0082 V
32.9999 V 30.0000 V 45 Hz 29.9957 V 30.0044 V
32.9999 V 30.0000 V 1 kHz 29.9957 V 30.0044 V
32.9999 V 30.0000 V 10 kHz 29.9957 V 30.0044 V
32.9999 V 30.0000 V 20 kHz 29.9928 V 30.0072 V
32.9999 V 30.0000 V 50 kHz 29.9904 V 30.0096 V
32.9999 V 30.0000 V 90 kHz 29.9759 V 30.0241 V
329.999 V 33.000 V 45 Hz 32.993 V 33.007 V
329.999 V 33.000 V 10 kHz 32.989 V 33.011 V
329.999 V 300.000 V 45 Hz 299.953 V 300.047 V
329.999 V 300.000 V 1 kHz 299.953 V 300.047 V
329.999 V 300.000 V 10 kHz 299.946 V 300.054 V
329.999 V 300.000 V 18 kHz 299.928 V 300.072 V
329.999 V 300.000 V 50 kHz 299.922 V 300.078 V
329.999 V 200.000 V 100 kHz 199.630 V 200.370 V
1020.00 V 330.00 V 45 Hz 329.91 V 330.09 V
1020.00 V 330.00 V 10 kHz 329.91 V 330.09 V
1020.00 V 1000.00 V 45 Hz 999.74 V 1000.26 V
1020.00 V 1000.00 V 1 kHz 999.79 V 1000.21 V
1020.00 V 1000.00 V 5 kHz 999.79 V 1000.21 V
1020.00 V 1000.00 V 8 kHz 999.74 V 1000.26 V
1020.00 V 1020.00 V 1 kHz 1019.79 V 1020.21 V
1020.00 V 1020.00 V 8 kHz 1019.74 V 1020.27 V
Note: Typical specification is -24 dB at 2 MHz