Calibration and Verification
Calibration
3
3-13
Table 3-9. Calibration Steps for DC Current
Step 5520A Output (AUX HI, LO) Shunt to Use
1 300.000 µA Fluke 742A-1k 1 k Resistance Standard
2 3.00000 mA Fluke 742A-100 100 Resistance Standard
3 30.000 mA Fluke 742A-10 10 Resistance Standard
4 300.000 mA Fluke 742A-1 1 Resistance Standard
5 2.00000 A Guildline 9230 0.01 shunt
20A, LO
6 10.0000 A Guildline 9230 0.01 shunt