5520A
Service Manual
3-16
Table 3-11. Calibration Steps for AC Current
5520A Output (AUX HI, LO)
Step
Amplitude Frequency Shunt to Use
1 3.29990 mA 100.00 Hz Fluke A40 10 mA
2 0.33000 mA 100.00 Hz Fluke A40 10 mA
3 3.00000 mA 10.00 kHz Fluke A40 10 mA
4 3.00000 mA 30.000 kHz Fluke A40 10 mA
5 0.30000 mA 100.00 Hz Fluke A40 10 mA
6 0.30000 mA 10.00 kHz Fluke A40 10 mA
7 0.30000 mA 30.00 kHz Fluke A40 10 mA
8 30.0000 mA 100.00 Hz Fluke A40 200 mA
9 30.0000 mA 10.00 kHz Fluke A40 200 mA
10 30.0000 mA 30.00 kHz Fluke A40 200 mA
11 300.000 mA 100.00 Hz Fluke A40 2 A
12 300.000 mA 10.00 kHz Fluke A40 2 A
13 300.000 mA 30.00 kHz Fluke A40 2 A
14 2.00000 A 100.00 Hz Fluke A40 2 A
15 2.00000 A 1000.0 Hz Fluke A40 2 A
16 2.00000 A 5000.0 Hz Fluke A40 2 A
17 2.00000 A 60.00 Hz Fluke A40 2 A
18 2.00000 A 100.00 Hz Fluke A40 2 A
19 2.00000 A 440.00 Hz Fluke A40 2 A
AUX 20A, LO
20 10.0000 A 100.00 Hz Fluke A40A 20 A
21 10.0000 A 500.00 Hz Fluke A40A 20 A
22 10.0000 A 1000.00 Hz Fluke A40A 20 A
23 10.0000 A 60.00 Hz Fluke A40A 20 A
24 10.0000 A 100.00 Hz Fluke A40A 20 A
25 10.0000 A 440.00 Hz Fluke A40A 20 A