Service Manual
3-30
CAL_INFO?
Description: Return message or instructions associated with running step
Response: (String) the message string
CAL_NEXT
Description: Continue a calibration procedure if it is waiting for a CAL_NEXT
command.
Parameter: (Optional) reference value (used if it’s waiting for a reference) If the
reference value has no unit, the unit is assumed to be that returned by the
CAL_REF? command
Example: CAL_NEXT
CAL_NEXT 2.999987
CAL_REF?
Description: Return nominal value expected for reference entry
Response: 1. The nominal value
2. The accepted or implied unit
3. Example: 3.000000e+00,V
CAL_SKIP
Description: Skip to next entry point in calibration procedure.
CAL_SECT
Description: Skip to next section of calibration procedure.
CAL_START
Description: Start a calibration procedure
Parameter: 1. Procedure name:
MAIN is the procedure for the 5520A minus any scope cal option
ZERO is the internal procedure to correct zero offsets
OHMSZERO is the internal procedure to touch up resistance offsets
SCOPE is the procedure for the 5520A-SC300 scope cal option
SC600 is the procedure for the 5520A-SC600 scope cal option
DIAG is the diagnostic pseudo-cal procedure
NOT aborts a procedure after the step underway