5520A
Service Manual
3-32
EOFSTR?
Description: Returns the End-Of-File character string used for calibration reports
Parameter: None
Response: (String) The End-Of-File character string
PR_RPT
Description: Prints a self-calibration report out the selected serial port
Parameter: 1. Type of report to print: STORED, ACTIVE, or CONSTS
2. Format of report: PRINT (designed to be read)
SPREAD (designed to be loaded into a spreadsheet)
3. Calibration interval to be used for instrument specifications in the
report: I90D (90 day specifications) or I1Y (1 year specifications)
4. Serial port out which to print report: HOST or UUT
Example: PR_RPT STORED,PRINT,I90D,HOST
RPT?
Description: Returns a self-calibration report.
Parameter: 1. Type of report to return: STORED, ACTIVE, or CONSTS
2. Format of report: PRINT (designed to be read)
SPREAD (designed to be loaded into a spreadsheet)
3. Calibration interval to be used for instrument specifications in the
report:
I90D (90 day specifications) or I1Y (1 year specifications)
Example: RPT? STORED,PRINT,I90D
RPT_PLEN
Description: Sets the page length used for calibration reports. This setting is stored in
nonvolatile memory.
Parameter: Page length
RPT_PLEN?
Description: Returns the page length used for calibration reports.
Parameter: None
Response: (Integer) Page length