EasyManua.ls Logo

Fluke 5520A - Page 93

Fluke 5520A
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration and Verification
Performance Verification Tests
3
3-37
Table 3-22. Verification Tests for DC Current (AUX Output)
Range Output Lower Limit Upper Limit
329.999 µA 0.000 µA -0.020 µA 0.020 µA
329.999 µA 190.000 µA 189.957 µA 190.043 µA
329.999 µA -190.000 µA -190.043 µA -189.957 µA
329.999 µA 329.000 µA 328.941 µA 329.059 µA
329.999 µA -329.000 µA -329.059 µA -328.941 µA
3.29999 mA 0.00000 mA -0.00005 mA 0.00005 mA
3.29999 mA 1.90000 mA 1.89980 mA 1.90020 mA
3.29999 mA -1.90000 mA -1.90020 mA -1.89980 mA
3.29999 mA 3.29000 mA 3.28969 mA 3.29031 mA
3.29999 mA -3.29000 mA -3.29031 mA -3.28969 mA
32.9999 mA 0.0000 mA -0.00025 mA 0.00025 mA
32.9999 mA 19.0000 mA 18.9982 mA 19.0018 mA
32.9999 mA -19.0000 mA -19.0018 mA -18.9982 mA
32.9999 mA 32.9000 mA 32.8971 mA 32.9029 mA
32.9999 mA -32.9000 mA -32.9029 mA -32.8971 mA
329.999 mA 0.000 mA -0.0025 mA 0.0025 mA
329.999 mA 190.000 mA 189.982 mA 190.018 mA
329.999 mA -190.000 mA -190.018 mA -189.982 mA
329.999 mA 329.000 mA 328.971 mA 329.029 mA
329.999 mA -329.000 mA -329.029 mA -328.971 mA
2.99999 A 0.00000 A -0.00004 A 0.00004 A
2.99999 A 1.09000 A 1.08979 A 1.09021 A
2.99999 A -1.09000 A -1.09021 A -1.08979 A
2.99999 A 2.99000 A 2.98906 A 2.99094 A
2.99999 A -2.99000 A -2.99094 A -2.98906 A
20.5000 A 0.0000 A -0.0005 A 0.0005 A
20.5000 A 10.9000 A 10.8954 A 10.9046 A
20.5000 A -10.9000 A -10.9046 A -10.8954 A
20.5000 A 20.0000 A 19.9833 A 20.0168 A
20.5000 A -20.0000 A -20.0168 A -19.9833 A

Table of Contents

Other manuals for Fluke 5520A

Related product manuals