6270A
Service Manual
4
The Product Manual Set
The Product manuals include:
• 6270A Safety Information (ships with the Product)
• 6270A Operators Manual (available on the Fluke Calibration)
• 6270A Remote Programmers Guide (available on the Fluke Calibration
website)
• 6270A Service Manual (available on the Fluke Calibration website)
To order, contact a Fluke Calibration sales representative. See Contact Fluke
Calibration.
Specifications
General Specifications
Power Requirements .......................... 100 V ac to 240 V ac, 47 Hz to 63 Hz
Fuse .................................................... T2A 250 V ac
Max Power Consumption .................... 100 W
Operating Ambient
Temperature Range ............................ 15 °C to 35 °C
Storage Temperature .......................... -20 °C to 70 °C
Relative Humidity
Operating ...................................... <80 % to 30 °C, <70 % to 40 °C
Storage ......................................... <95 %, non-condensing. A power stabilization period of four days may
be required after extended storage at high temperature and humidity.
Vibration .............................................. MIL-T-28800E
Altitude (Operation) ............................. <2000 m
Ingress Protection ............................... IEC 60529: IP20
Safety .................................................. IEC 61010-1, Installation Category II, Pollution degree 2
Warmup Time ..................................... 15 minutes after power up or module installation, when items previously stored
within Operating Ambient Temperature Range.