Pressure Controller/Calibrator
Module Guide Rails
35
PMM Dock Assembly
Use the PMM dock assembly to mount the PM600 and PM200 transducer
modules.
 Warning
To prevent possible electrical shock, fire or personal injury,
turn off the Product and disconnect the mains power cord prior
to the removal procedure. Shut off the supply pressure and
vent the system before removing the manifold, valve, or
associated parts.
To remove the PMM dock assembly or associated parts, see Figures 17, 18, and 19:
1. Make sure the Product is off and disconnect the mains power cord.
2. Shut off supply pressure and vent the system.
3. Remove all pressure lines and fittings from the manifold if necessary to move
equipment.
4. Remove the PCM and PMM modules from all slots. See PMM and PCM
Removal.
5. Remove the test manifold.
6. Remove top panel, see Top Panel.
7. Remove the COMM Tray, see Communications Tray.
8. Open the module bay. Pull the latch to remove the tab of the door frame
that blocks the door open sensor .
9. Disconnect the power cable from the PCM solenoid driver PCA J1 .