Pressure Controller/Calibrator
Power Supply and Fan
45
2
1
3
4
5
6
8
7
hxn227.eps
Figure 24. Remove the G-Series Valve
Power Supply and Fan
To replace the power supply, fan, or associated parts, use the subsequent
instructions and Figures 25 and 26. The power supply and fan parts are listed in
Table 49.
Note
After any portion of the ac power system is serviced, perform
appropriate Hypot and Ground Bond testing before you put the
Product back into service.
Warning
To prevent possible electrical shock, fire or personal injury,
turn off the Product and disconnect the mains power cord prior
to the removal procedure. Shut off the supply pressure and
vent the system before you remove the manifold, valve, or
associate parts.