6270A
Service Manual
102
Table 32. PM200-G20M Pressure Point Matrix
As Found / As Left Points
Calibration Segment Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging Time
[s]
Ascending
Points
1 0
1
60 30
2 4000 60 30
3 8000 60 30
4 12 000 60 30
5 16 000 60 30
6 20 000 60 30
Descending
Points
7 16 000 60 30
8 12 000 60 30
9 8 000 60 30
10 4 000 60 30
11 0 60 30
1. UUT is zeroed before setting pressure.
PM600 Test Definition Matrices
Table 33. PM600-BG15K Pressure Point Matrix
As Found / As Left Points
Calibration Segment Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging Time
[s]
Ascending Points
1 -15
1
90 30
2 -7.5 90 30
3 -0.015 90 30
4 7.5 90 30
5 15 90 30
Descending Points
6 7.5 90 30
7 0.015 90 30
8 -7.5 90 30
9 -15 90 30
1. UUT is zeroed before setting pressure.